Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 36: (1 year) CAL VER /5520 DATE: 05-Mar-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 37 NUMBER OF LINES: 142 CONFIGURATION: Fluke 5520A STANDARD: Fluke 5500A/COIL ============================================================================= # # Source: # Fluke 36 AC/DC True RMS Clamp Meter Service Information # PN 200409, September 1996 # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later # # Subprocedures: # None # # Required Files: # 52coil3x.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5520A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 STD Fluke 5500A/COIL 1.002 ASK- P F W 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP Ensure that all test equipment has been allowed to 1.006 DISP stablize in an environment of 23 +/-5degC and that all 1.006 DISP specified equipment warm-up times have been allowed to 1.006 DISP elapse. 1.007 HEAD {DC VOLTAGE} 1.008 DISP Connect the 5520A to the UUT as follows: 1.008 DISP [32] 5520A NORMAL Output Hi to UUT V-Ohm 1.008 DISP [32] 5520A NORMAL Output Lo to UUT COM 1.009 HEAD DC VOLTAGE: {200V Range} 1.010 5520 200 20.0V 0.5U 2W 2.001 5520 200 100.0V 1.3U 2W 3.001 5520 200 -200.0V 2.3U 2W 4.001 HEAD DC VOLTAGE: {600V Range} 4.002 5520 600 -600V 9U 2W 5.001 HEAD {AC VOLTAGE} 5.002 HEAD AC VOLTAGE: {200V Range} 5.003 5520 200 200.0V 3.3U 50H SI 2W 6.001 5520 200 200.0V 3.3U 60H SI 2W 7.001 5520 200 200.0V 3.3U 400H SI 2W 8.001 HEAD AC VOLTAGE: {600V Range} 8.002 5520 600 100V 5U 50H SI 2W 9.001 5520 600 100V 5U 60H SI 2W 10.001 5520 600 100V 5U 400H SI 2W 11.001 5520 100V 60H SI S 2W 11.002 DISP Press the MAX button. 11.003 5520 * S 11.004 MEMI Enter UUT reading in volts: 11.005 MEME 11.006 MEMC 600 V TOL 15V 12.001 5520 600 600V 12U 50H SI 2W 13.001 5520 600 600V 12U 60H SI 2W 14.001 5520 600 600V 12U 400H SI 2W 15.001 HEAD {RESISTANCE} 15.002 DISP Connect the 5520A to the UUT as follows: 15.002 DISP [32] 5520A AUX Output Hi to UUT V-Ohm 15.002 DISP [32] 5520A AUX Output Lo to UUT COM 15.002 DISP [32] 5520A NORMAL Output Hi to UUT V-Ohm 15.002 DISP [32] 5520A NORMAL Output Lo to UUT COM 15.003 HEAD RESISTANCE: {200 Ohm Range} 15.004 RNG 200Z 15.005 5520 30.0Z S CW 15.006 EVAL -s "Beeper on" : Is the beeper on? 16.001 5520 70.0Z S CW 16.002 EVAL -s "Beeper off" : Is the beeper off? 17.001 5520 200 70.0Z 1.3U CW 18.001 5520 200 190.0Z 2.8U CW 19.001 DISP Disconnect the 5520A from the UUT. 19.002 HEAD {DC CURRENT} 19.003 HEAD DC CURRENT: {Zero} 19.004 RNG 200A 19.005 DISP Turn DC Amps ZERO knob fully clockwise. 19.006 EVAL -s "3.0 to 19.0" : Does UUT indicate between 3.0 and 19.0? 20.001 DISP Turn DC Amps ZERO knob fully counter-clockwise. 20.002 EVAL -s "-19.0 to -3.0" : Does UUT indicate between -19.0 and -3.0? 21.001 HEAD DC CURRENT: {200A Range} 21.002 PIC 52coil3x 21.003 5520 200 20.0A 1.1U T5 22.001 5520 200 -50.0A 1.7U T5 23.001 DISP Move the Hi lead from the 5520A AUX Hi to the 20A. 23.002 5520 200 200.0A 4.5U HC T5 24.001 HEAD DC CURRENT: {Balance} 24.002 5520 200.0A HC S T5 24.003 DISP Move conductor from center of jaw to top. 24.004 MEMI Enter UUT reading in amps: 24.005 MEME 24.006 DISP Move conductor from top of jaw to bottom. 24.007 MEMI Enter UUT reading in amps: 24.008 MEME 24.009 MEMC 200 A 2.0U 25.001 HEAD DC CURRENT: {1000A Range} 25.002 DISP [32] WARNING 25.002 DISP Do not sustain any of the 1000A Range tests for more 25.002 DISP than 2 minutes or the 5520A/COIL may be damaged. 25.003 5520 1000 600A 15U HC T5 26.001 5520 1000 -600A 15U HC T5 27.001 5520 1000 900A 30U HC T5 28.001 5520 1000 -900A 30U HC T5 29.001 HEAD {AC CURRENT} 29.002 DISP Move the Hi lead from the 5520A 20A to AUX Hi. 29.003 HEAD AC CURRENT: {200A Range} 29.004 5520 200 20.0A 0.9U 50H SI LT5 30.001 5520 200 20.0A 0.9U 60H SI LT5 31.001 5520 200 20.0A 0.9U 400H SI LT5 32.001 DISP Move the Hi lead from the 5520A AUX Hi to the 20A. 32.002 5520 200 200.0A 4.5U 50H SI HC LT5 33.001 5520 200 200.0A 4.5U 60H SI HC LT5 34.001 5520 200 200.0A 4.5U 400H SI HC LT5 35.001 HEAD AC CURRENT: {600A Range} 35.002 5520 600 600A 15U 50H SI HC LT5 36.001 5520 600 600A 15U 60H SI HC LT5 37.001 END